European Norm EN554 - Sterilization of medical devices - Validation and routine checking of steam sterilizers - is the reference standard for operational qualification of steam autoclaves.
The norm demands that :
- Temperature be measured at the core of the load
- Saturated steam be checked at base plate level
- Temperature and pressure be maintained constant while on hold pattern
- Sterilization temperature while on hold pattern be :
- within specified temperature range, the upper control limit being equal to sterilization temperature plus 3°C,
- with no deviation exceeding 1°C,
- and temperature difference of any 2 points inferior to 2°C.
- Temperature be set within 15s in autoclave whose volume is inferior to 800L, and within 30s in more voluminous ones.
ValiVACQ system is TMI-ORION's response to the demand
The loggers can be located at the core of loads. They log temperature and pressure versus time.
 Contrôle du cycle de stérilisation au coeur d'une charge d'instruments | Contrôle du cycle de stérilisation au coeur d'une charge de linge |
Valivacq system performances meets or exceeds the recommendations of the guide of European standard EN554, published by AFNOR (Reference GA S 98130).
- Temperature accuracy better or equal to 0.17°C - Valivacq: 0.10°C
- Pressure accuracy better or equal to16mbar - Valivacq: 10 mbar
Furthermore, the Valivacq software makes it possible to publish a report in compliance with norm EN554 requirements. (see ValiVACQ data sheet in "data processing software").
Routine control of steam sterilizers
Thanks to an excellent reliability at elevated temperature, the TMI-Orion data loggers are very suitable to the routine control of the autoclaves specifications during sterilization cycles.
Ethylen Oxide sterilization control top
For ethylen oxide low temperature sterilization process the NanoVACQ-Ex data loggers have temperature and humidity/temperature sensors.
These loggers can be programmed and read out with multi-logger interface. They are designed according the Ex norms (explosion proof).
Washing and desinfection temperature control

NanoVACQ Flat MiniVACQ
Two data logger models fit the control of temperature for washing and desinfection processes: NanoVACQ Flat and MiniVACQ. They can easily be placed inside the equipments thanks to their small dimensions.